Reference/Railing Artifact
Note: A reference/railing artifact will display an overly amplified signal, the data will appear to "rail" vertically on one or all channels.
 Reference Issue
To Resolve: First, observe the disposition of L and R channels.
Check for:
-
Is the back of the sensor touching the patient's ear?
-
Is the hair swept away and back from the sensor?
Is the HydroDot pressed into the socket far enough?
Next:
|
 Recheck> L
 >The test session will pause and prompt a recheck of L or R.
|
|